Que. : What are the different methods of temperature
measurement ? Explain.
Ans. : The different methods of temperature
measurement are :
1. MECHANICAL 2. ELECTRICAL.
MECHANICAL
METHOD :
1. Mercury in
glass thermometers : This consists of a glass tube of very fine bore joined to
a reservoir at the bottom and sealed at the top. A measured quantity of mercury
is the enclosed. When the thermometer is heated the mercury expands much more
than the glass and is therefore forced to rise up in the tubing A scale is
fixed at the side.
2. Bimetallic
Thermometer : Two metals whose coefficient of linear expansion is different are
welded and rolled together to the desire thickness. The actual movement of a
bimetal is its flexivity with one end fixed, a straight bimetal strip deflects
in proportion to its temperature, to the square of its length and inversely
with its thickens.
Pressure
Spring Thermometers :
There are four
classes of pressure spring thermometers.
1. Liquid
filled.
2. Vapor
pressure.
3. Gas filled.
4. mercury
filled.
Liquid
filled & Mercury filled :
Both type,
operate on the principle of thermal expansion. Where the bulb is immersed in a
heated substance. The liquid expands causing the pressure spring to unwind. The
indicating, recording or controlling mechanisms are attached to pressure
spring.
Compensated
Thermometer System :
Compensations
are provided in order to nullify the effect of changes in ambient temperature.
The compensation in liquid filled expansions thermal system consists of the
second tubing and helical element, both liquid filled. The two elements are so
constructed that the measuring helical floats on a movable base the position of
which is governed by the compensating helical. The two tubing and helicals are
matched in volume so that variation in temperature at the instrument case and
along the capillary tubing produce equal motion from both helicals. Such motion
nullity each other so that only motion produced by varying the bulb temperature
actuates the recorder pen.
Gas filled
Thermometers :
This type
depends upon the increase in pressure of a confirm gas (constant volume) due to
temp. increase. The relate between temp. and pressure in this kind of system
follow Charles law and may be expressed.
P1
P2
The system is
filled under high pressure. The increase pressure for each degree of
temperature rise is therefore greater than if the filling pressure were low.
Nitrogen the gas most after used for such systems, because it chemically insert
and possesses a favorable coefficient thermal expansion.
Vapor -
Pressure Thermometers :
Vapor pressure
thermometers depend upon vapor pressure of liquid which only partially fills
the system. At low temperatures the vapor pressure increase for each unit
temperature charge is small, at higher temperature the vapor pressure change is
much greater.
ELECTRICAL
METHOD OF TEMPERATURE MEASUREMENT :
1.
THERMOCOUPLES : It is a simple device consisting of a dissimilar metal wires
joined at their ends. when an of each wire is connected to a measuring
instrument thermocouples becomes an accurate and sensitive temperature
measuring device.
Type of
Thermocouples :
1. Iron -
Constantan (J) ---- 0 - 800'C.
2. Chromel
Alumel ( K ) ---- 0 - 1000'C.
3. Platinum -
Platinum % Rhodium ---- 0 - 1500'C.
4. Platinum -
Rhodium 13% (S) ---- 0 - 1700'C.
5. Chromel -
Constantan (E) ---- 0 -700'C.
6. Copper -
Constantan (T) ---- 0 - 400'C.
RESISTANCE
- TEMPERATURE DETECTERS (RTD) :
RTD's are
generally used for precise temperature measurement. It consists of a five wire
wrapped around an insulator and enclosed in a metal. The most sheath of a
resistance thermometer resembles that of bimetallic thermometer bulb.
PRINCIPLE :
"Resistance increases as temperature increase"
Rt. = Ro (1 +
xt)
Rt. =
Resistance of Temperature to measured.
Ro. =
Resistance of zero temperature.
x = Co. off of thermal (expansion).
t = Temperature to be measured.
These metals
have a positive temperature co-efficient of expansion. Therefore resistance
increases as the temperature increases.
CALCULATION
OF RESISTANCE FOR Pt100.
Ro.
= 100
x for
platinum = 0.00385 /c.
To calculate
Resistance at 100'c.
R100 = 100 [ 1+ ( 38.5 x 10
4 x 100 ) ]
= 100 + (100 x
0.385)
R100 = 138.5
Resistance at
100'c = 138.5
Que. : What is Pt 100 mean. ?
Ans. : Pt100 means 100 OHMS at 0'C for a platinum
resistance bulb.
Que. : What is two wire and three wire R.T.D. system
?
Ans. : Two wire R .T .D. system :
Two wire RTD
system use for short distance like a compressor field local panel.
Three wire
System :
Three wire
system use for long distance coke a field to control Run.
The third wire
is used for compensation of lead wire resistance.
Que. :Draw a
potentiometric temp. measuring circuits and explain its . ?
Ans. :
OPERATION :
The input to
the instrument is a measurement of some in the processes using a sensing
element ( such as thermocol ) or a device to produce direct voltage, which is
the voltage (signal). This voltage is subtracted from a voltage developed by a
known constant voltage in a potentiometric measuring circuit. The subtraction
occurs by connecting two voltages in series with the opposing polarity,
difference between these two voltages produces signal, the voltage going to the
amplifier. The error will positive or negative depending on which of the two
voltages greater. When amplified, the error signal will drive servo balancing
motor in appropriate direction to adjust circuit ( actually drive the slide
wire ) until the difference between the feedback voltage and the input voltage
is balance out. An error signal equal to zero results ( null point ) the
balancing ( servo motor is be longer driven )
Que. : What is the constant voltage unit ?
Ans. : The constant voltage circuit consists of a
rectifier, CR, a filter capacitor C1, followed by two stages of zener regulation.
Abridge configuration is provided to 1amp line voltage regulation zener CR3, R1
and R2 combine provide relatively constant current to zener CR4, Thus
variations. Resisters R2 and R3 form a bridge that any remoment line voltage
effects.
Que. : Explain the working of a balancing motor.
Ans. : Signal in control win appears as due tank
circuit formed by win and capacitor of
amp. board.
+ve signal :
It lags 90' from due to line phase capacitor amp. board.
-ve
signal : Leads 90' from line due to line
phase capacitor of amp. board.
The servo (
balancing ) motor is an induction motor that functions by creating a rotating
magnetic field in the stator.
The rotor (
armature ) turns by following this field. The field is developed by the use of
two windings in the stator.
It has got two
windings, one of which is continuously energized by the line voltage. The other
winding is energized by the power amplifier, with a current whose phase with
respect to line current determines the direction of rotation of motor.
Que. : What is burnout feature ? Explain.
Ans. : Burnout provides the warning feature of
driving indicator the end of scale if the input circuit should open.
A burnout
resistor is provided which develops a voltage drop between the measuring
circuit and the amplifier. The polarity of the signal determines the direction
of the servo drive upon an open circuit in the input.
Upscale
burnout : R value 10 M
Downscale
burnout : R value 2.2 M
Que. : Why is
a converter used in a temp. recorder ?
Ans. : The
converter is designed to convert D. C. input voltage into an A. C. input
voltage proportional in amplitude to the input.
Que. : Why
are thermowells used ?
Ans. : In
numerous application it is neither desirable nor practical to expose a
temperature sensor directly to a process material. Wells are therefore used to
protect against damage corrosion, errosion, abrasion and high pressure
processes. A thermowell is also useful in protecting a sensor from physical
damage during handling and normal operation.
Selecting a thermowell :
The significant properties considered in selecting a
material for the well are as follows :
1. Resistance to corrosion and oxidation.
2. Resistance to mechanical and thermal shock.
3. Low
permeability ( Resistance to gas leakage ).
4. Mechanical
strength.
5. Thermal
conductivity.
Material
for Wells :
1. Stainless
steal.
2. Inconel.
3. Monel.
4. Alloy
steal.
5. Hastelloy
'C'.
Que. : How will you calibrate a temp. recorder using
a potentiometer ?
Ans. : Connect the potentiometer output the input of
temp. recorder.
1. Connect the
( +ve ) to the ( +ve ) and ( -ve ) to the ( -ve ).
2. If ambient
compensation is provided in potentiometer set it to the correct ambient temp.
3. If no
ambient compensation is provided take a thermometer and measure the correct
ambient temp.. Find out the corresponding mV s for that temp. for the given
input type of thermocouple.
4. While
feeding subtract the ambient temp. m v s from the corresponding temp. every
time.
5. Adjust the
necessary adjustments.
MEASURING
TEMPERATURE WITH A POTENTIOMETER :
1. Connect the
input of the potentiometer to the thermocouple.
2. If no
ambient compensation is provided find out the corresponding millivolts for that
ambient temp. for the type of thermocouple used.
3. Add the
ambient temp. millivolts to the corr. input millivolts measured. Find out from
the chart the corresponding temperature.
Que. : What type of sensing element would you use to
measure very low temperature ?
Ans. : The sensing element used for measuring very
low temperature is R. T. D.
Que. : What are skin temperature thermocouples ?
Ans. : Skin thermocouples are those which are
directly connected to the process without any thermowell.
Used for measuring the skin
temperature of heaters furnaces, flue gas etc.
Que. : What is the specialty of thermocouples lead
wires ?
Ans. : They should be of the same material as the
thermocouple.
Que. : What is the difference the a wheatstone bridge
and a potentiometer ?
Ans. : The difference between a potentiometer and a
wheatstone bridge measuring instrument is that potentiometer is a voltage
measuring instrument and wheatstone bridge is a current measuring instrument.
Que. : Explain the continuos balance potentiometer
system using R. T. D.'s.
Ans. : In a balance wheatstone bridge resistance
thermometer a resistance bulb is connected into one branch of a d.c. bridge
circuit; in another branch is a variable resistance in the form of a calibrated
slidewire. Variations in temp. of the measured medium cause a change in
resistance of the bulb and a consequent unbalance of the bridge circuit. A self
balancing wheatstone bridge recognizes the condition of unbalance, determines
its direction and magnitude and position the slidewire contractor to rebalance
the bridge and indicate the temp. on the
scale.
The D. C.
potential appearing at AA is converted by the converting stage to an A. C.
voltage appearing at BB and is multiplied by the voltage amplifier to a large
value at cc. It is then used to control the power amplifier output DD which
drives the balancing motor in the proper direction to balance the bridge.
The polarity
of the signal at AA determines the phase of the alternating voltage at BB which
in turn determines the direction of rotation of the balancing motor.
Que. : How is automatic Reference junction
compensation carried out in temp. ? Recorders etc,.
Ans. : For automatic reference junction compensation a variable nickel resister is used. As the temperature changes, so does its resistance. This reference junction compensatory is located, so that it will be at the temperature of the reference junction. The reference junction is at the position where the dissimilar wire of the thermocouple is rejoined, which invariably is at the terminal strip of the instrument
No comments:
Post a Comment